Defibrilator ME PAD polo-aický
ME PAD is semi- aic defibrillator to help during sudden cardiac
death.
5 Years Warranty
Recording of up to 5... čítať viac...
The
ME PAD is semi- aic defibrillator to help during sudden cardiac
death.
5 Years Warranty
Recording of up to 5 operations of 3 hours duration each
on internal SD card
Including pads for adults
(up to 30 months storability)
Including lithium battery (up to 5 years storability)
Transportation bag and pads for pediatric (optional)
Fulfills military standarad MIL-STD-810G
Automatic volume adjustment to the surrounding volume
Emergency change to pediatric use (below 25kg)
without changing pads
Technické špecifikácie
ME PAD can be used at any place. The user gets all the necessary assistance as a voice instruction whereas voice volume is aically adjusted to the ambient noise.
Up to five operations of three hours of duration can be recorded on the internal SD memory card. At any time all the important data such as heart rhythm, energy output, time, etc. are available for the post-analysis.
In addition to the aic power switch while using children pads, the output can be lowered manually. Thus children can be treated in emergency cases even if pediatric electrodes are not available. In daily, weekly and monthly intervals a test program is performed.
All functions including battery performance and durability of the electrodes are aically checked. An alarm will be given in case of errors.
The software of ME PAD is programmed according to the current guidelines of the ERC.
The user is instructed through the entire resuscitation.
The long-life Li-ion battery guarantees a standby time of up to five years or up to 200 shocks at full power.
Technické špecifikácie
Output Energy Adult - 150 Joule at 50 Ω
Pediatric - 50 Joule at 50 Ω (common usage)
Charging Time less than 8 seconds
Charging Time after CPR finished at least 8 seconds
Waveform E-cube two-phase (truncated expotential type)
Acquired ECG Lead Lead II
Frequency Response 1 Hz to 30 Hz
Impendance Range 25 Ω to 175 Ω
(shock will not be delivered if the patient’s impedance is beyond this range)
Shockable Rhythms Ventricular Fibrillation or Fast Ventricular
Fibrillation
Sensitivity and Specificity
meets ANSI/AAMI DF80 guidelines
Control Devices Power Button, i-Button, Shock Button, Adult/
Pediatric Selection Switch
Status LCD Displays device status, battery level and pads
status
Speaker Plays back voice instructions. The CU-SP1
analyzes the ambient noise level during a treatment operation. If ambient noise level is
high, it aically increases the voice instructions volume so that you can hear them
clearly.
Auto Power On Self-Test, Run-time Self-Test
Daily, Weekly, and Monthly Self-Test
Manual
Battery Pack Insertion Test (done when the user
inserts the battery pack into the battery pack compartment of the device)
Disposable Battery Pack
Battery Type 12V DC, 4.2Ah LiMnO2, Disposable: Long-life
Capacity
At least 200 shocks for a new battery or 8 hours of operating time at room temperature
Standby Life
(After inserting battery)
At least 5 years from the date of manufacture if stored and maintained in accordance with the
instructions in this document.
Temperature Ranges
Operating: 0° ~ 43° C
Storage: -20° ~ 60° C
Adult Defibrillation Pads
Electrode Area: 120 cm²
Cable Length: Total 120 cm (Inside the pouch:95 cm, Outside the pouch: 25 cm)
Shelf life: Up to 30 months from the date of manufacture
Pediatric Defibrillation
Pads
Electrode Area: 46,43 cm²
Cable Length: Total 120 cm (Inside the pouch: 80 cm, Outside the pouch: 40 cm)
Shelf life: Up to 24 months from the date of manufacture
Internal Memory Data
Capacity 5 individual treatments, up to 3 hours per treatment
SD Card
External memory. Data may be copied from the internal memory to the SD Card
IrDA For PC communications
Sealing meets DIN EN 60529: IP55
ESD meets EN 61000-4-2:2001
EMI (Radiated) meets EN 60601-1-2 limits, method EN
55011:2007 + A2:2007, Group 1, Class B
EMI (Immunity)
meets IEC 60601-1-2 limits, method EN 61000-
4-3:2006 +A1:2008 Level 3
(10V/m 80MHz to 2500MHz)
Vibration Operating: Meets MIL-STD-810G Fig.514.6E-1,
random
Standby: Meets MIL-STD-810G Fig.514.6E-2,
swept sine(helicopter)
Altitude 0 to 15,000 feet (operational and storage)
Drop withstands 1.2-meter drop to any edge, corner, or surface
Environmental
Conditions
Operation: 0° ~ 43° C, 5% ~ 95% (non
condensing)
Standby: 0° ~ 43° C, 5% ~ 95% (non
condensing)
Transport: -20° ~ 60° C, 5% ~ 95% (non
condensing), device only
Languages German, English, French, Danish, Spanish,
Norwegian, Italian, Czech, Portuguese, Greek,
Swedish, Dutch
More languages are planned
Dimensions 260 x 256 x 69,5 mm (WxLxH)
Weight 2,4 kg including battery pack & pads